					Oxygensensor part nr 
					(定货型号) | 
					Output signal in air (空气中输出)
 | 
					Response  time T 90% (响应时间)
 | 
					Nominal sensor lifetime (使用寿命)
 | 
					Electrical interface  (电路接口类型)
 | 主要兼容品牌/型号/代码 | 
				
					|   | 
					
					OOM101 |  46µA-63µA no temperature compensation |  <12 seconds |   > 500 000 % O2 
					hours |  gold plated slip rings | Analytical Industries 
					PSR-11-915 | 
				
					| Biotest S101 | 
				
					| Catalyst Research MSA 
					(梅思安) 655264 | 
				
					| Ceramatec CAG-1 | 
				
					| Draeger 6803290 | 
				
					| Hamilton (夏美顿) HM-10 | 
				
					| Hewlett-Packard (HP惠普) 
					15201-1A | 
				
					| Hudson 5556 | 
				
					| IT-Gambert M-01 | 
				
					| Maxtec (迈克思泰克) 
					MAX-1、MAX-01 | 
				
					| Datex Ohmeda (欧美达) 
					4700 Oxicap,5250 RGM,Excel Modulus, 5120 Monitor,5125 
					Monitor,5140,7200,3000/5, 5250,5100,5400, 7800/7810 
					Vent.Aespira, Avance, Aisys,7800,7810. 0237-2034-700 | 
				
					| Pacifitech PT-10 | 
				
					| Paragon Services 
					PS-1000 | 
				
					| Sensor Tec (BPR 
					Medical) ST-10 | 
				
					| Sensidyne SV-14A | 
				
					| Teledyne C-41131 (R15) | 
				
					| Nuova E-15/0 | 
				
					| VTI (Vascular 
					Technology ) VT1272, 101272 100352 | 
				
					|   | 
					
					OOM102 |  9mV-14mV temperature compensated | <12 seconds  |  > 1 000 000 % O2 
					hours | 3 pin Molex connector
					 | Airshields Isolette (Hill Rom) C2000/C200 8362030 | 
				
					| ATOM 
					2100 Incubator 、2000 | 
				
					| Biotest A103 | 
				
					| City 
					Technology (城市技术) MOX-1 、MOX-01 | 
				
					| Criticare (8100 Poet, 8500 Poet) | 
				
					| CSI 
					644 cell | 
				
					| Datascope 0600-00-0002 | 
				
					| Sensor Tec (BPR Medical) ST-12 | 
				
					| Space Labs 90514 | 
				
					| Teledyne, Hill-ROM,GE Marquette R22 MED | 
				
					| Maxtec (迈克思泰克) MAX-12A | 
				
					| Ohmeda Centiva S/5 | 
				
					| Nuova E-17/A2 | 
				
					|   | 
					
					
					OOM102-1 | 9mV-14mV temperature compensated |  <12 seconds |  > 1 000 000 % O2 
					hours | 3.5mm mono Jack 
					 | Analytical Industries 
					PSR-11-917-J | 
				
					| ATOM OX-60 | 
				
					| BMD Bio Med Devices 
					M-10, M-2 10006 | 
				
					| Bird Products (VIASYS) 
					(鸟牌) 6400 Vent., 8400 Vent., VIP Vent. | 
				
					| Catalyst Research MSA 
					(梅思安) MiniOX I,II,III & 3000 406931 | 
				
					| City Technology (城市技术) 
					MOX-2 、MOX-02 | 
				
					| Dameca Alpha Med 
					42035-46 | 
				
					| Datex-Engstrom  Erica 
					、Elvira | 
				
					| Hudson 5567 | 
				
					| Puritan Bennett 
					(PB泰科)7280 Monitor 7200 Ventilator 4-020933-00 | 
				
					| Sensor Tec (BPR 
					Medical) ST-11 | 
				
					| Stephan Artec | 
				
					| Taema (天马) Taema Alys, 
					Claris YR010100 | 
				
					| Teledyne TED-AX300 
					(R17 MED) | 
				
					| MSA MiniOx 3000 | 
				
					| Nuova E-15/5 | 
				
					|   | 
					
					OOM103 |  9mV-13mV temperature compensated |  <5 seconds |  > 500 000 % O2 
					hours | 3 pin Molex connector
					 | Criticare Poet | 
				
					| Maxtec (迈克思泰克) MAX-22CC | 
				
					| Teledyne R22CC | 
				
					| Nuova E-17/A2 | 
				
					|   | 
					
					
					OOM103-1 |  9mV-13mV temperature compensated |  <5 seconds |  > 500 000 % O2 
					hours | 3.5mm mono Jack 
					 | Invivo 9445 | 
				
					| Maxtec (迈克思泰克) MAX-13F | 
				
					| Nuova E-17/J | 
				
					|   | 
					
					
					OOM103-1M |  9mV-13mV temperature compensated | <5 seconds  |  > 500 000 % O2 
					hours | Switchcraft mini power Jack
					 | IT-Gambert M-25 | 
				
					| Maxtec (迈克思泰克) MAX-250esf | 
				
					|   | 
					
					OOM104 |  24µA-32µA no temperature compensation | <12 seconds  |  > 750 000 % O2 
					hours | gold plated slip rings
					 | Drager (德尔格) Incubator 
					8000 6850645 | 
				
					| IT-Gambert M-08 | 
				
					| Sensor Tec (BPR 
					Medical) ST-13 | 
				
					| Nuova E-8000 | 
				
					|   | 
					
					OOM105 |  Teledyne TED range | <5 seconds  |  > 500 000 % O2 
					hours | Molex plug 4P4C 
					 | Analytical Industries PSR-11-77 | 
				
					| Catalyst Research MSA (梅思安) 804674 | 
				
					| City 
					Technology (城市技术) MOX-4 、MOX-04 | 
				
					| Hamilton (夏美顿) HM-03 | 
				
					| Hudson 5558 | 
				
					| IT-Gambert M-07 | 
				
					| Newport (纽邦) OM100 | 
				
					| Maxtec (迈克思泰克) MAX-17 | 
				
					| Sechrist 650 / 4370 | 
				
					| Sensor Tec (BPR Medical) ST-03 | 
				
					| Teledyne TED 60T/191T/200T T7 | 
				
					| Nuova E-77/0 | 
				
					|   | 
					
					OOM106 |  9mV-13mV temperature compensated | <12 seconds  |  > 1 000 000 % O2 
					hours |  3 pin Molex connector | Datascope 0600-00-0002 | 
				
					| Maxtec (迈克思泰克) MAX-12F | 
				
					| Nuova O-17/M1 | 
				
					| 
					
					
					 | 
					
					OOM107 |  170µA-230µA no temperature compensation | <12 seconds  |  > 250 000 % O2 
					hours | gold plated slip rings
					 | Catalyst Research MSA (梅思安) 485905 | 
				
					| Hamilton (夏美顿) Arabella, Amadeus, Veolar HM-01 | 
				
					| Hudson 5569 | 
				
					| IT-Gambert 
					M-05 | 
				
					| Maxtec (迈克思泰克) MAX-19 | 
				
					| Megamed M2 | 
				
					| Pacifitech PT-01 | 
				
					| Sensor Tec (BPR Medical) ST-01 | 
				
					| Teledyne No leads T1/T2 | 
				
					| Nuova E-33/0 | 
				
					|   | 
					
					
					OOM107-2 |  170µA-230µA no temperature compensation | <12 seconds  |  > 250 000 % O2 
					hours | flying leads with 
					pin-connectors  | Analytical Industries 
					PSR-11-33 | 
				
					| Bio Tek (宝待) 48351 | 
				
					| Catalyst Research MSA 
					(梅思安) 478841 | 
				
					| Hudson 5568 | 
				
					| IT-Gambert M-06 | 
				
					| Maxtec (迈克思泰克) MAX-18 | 
				
					| Megamed M1 | 
				
					| Pacifitech PT-02 | 
				
					| Sensor Tec (BPR 
					Medical) ST-02 | 
				
					| Teledyne With leads 
					C1R/ C2R | 
				
					| Vickers Mie | 
				
					| Nuova E-33/1 | 
				
					|   | 
					
					OOM110 |  10mV-12mV temperature compensated | <12 seconds  |  > 1 000 000 % O2 
					hours | modular Jack 6P4C | City 
					Technology (城市技术) MOX-7 、MOX-07 | 
				
					| IT-Gambert M-10 | 
				
					| Maxtec (迈克思泰克) MAX-10 | 
				
					| Ohmeda (欧美达) Aestiva, 7900 Monitor, Excel SE 6050-0004-110 | 
				
					|   | 
					
					OOM201 |  24µA-32µA (dual cathode) no temperature compensation | <12 seconds  |  > 500 000 % O2 
					hours | gold plated slip rings
					 | Analytical Industries 
					PSR-11-915-2 | 
				
					| Catalyst Research MSA 
					(梅思安) 655263 | 
				
					| Ceramatec CAG-11 | 
				
					| City Technology (城市技术) 
					MOX-16 | 
				
					| DP Medical MVA210 | 
				
					| Drager (德尔格) Cicero, 
					Cato, Babylog, Sulla, Oxycom, Oxydig, Evita 2 & 4, Anemone, 
					Evita Dura, Cicero, Babylog, Fabian, PM8030,Savina, Anemone, 
					Incubator 8000 , I-8000, 8000IC, Fabius GS, Narcomed 
					1A,2,2B,3,4   6850645 | 
				
					| NAD Narcomed 6850645 | 
				
					| Hamilton (夏美顿) 
					HM-13、HM-14 | 
				
					| Hewlett Packard (HP惠普) 
					15201A | 
				
					| Hudson 5557 , V-14 | 
				
					| IT-Gambert M-02 | 
				
					| Maxtec (迈克思泰克) MAX-11 | 
				
					| Mercury Medical 
					10-103-07 | 
				
					| Ohmeda O2 cell 
					6051-0000-224 | 
				
					| Oxitron 3710 | 
				
					| Pacifitech PT-14 | 
				
					| Paragon Services 
					PS-2000 | 
				
					| PPG Biomedical, Irisa 
					505055-001 | 
				
					| Sensor Tec (BPR 
					Medical) ST-14 | 
				
					| Sensidyne SV-14A | 
				
					| Teledyne R23 , R23MED | 
				
					| Vascular Technology 
					101372 | 
				
					| Ventrex V-14A | 
				
					| Nuova E-15/2 | 
				
					|   | 
					
					OOM202 |  13mV-16mV temperature compensated | <12 seconds  |  > 1 000 000 % O2 
					hours | 3 pin Molex  | Analytical Industries PSR-11-917-M | 
				
					| Bird 
					Products (VIASYS) (鸟牌) Avea. Vent | 
				
					| Catalyst Research MSA (梅思安) 472062 , 470644 | 
				
					| Ceramatec CAG-12 | 
				
					| City 
					Technology (城市技术) MOX-9 、MOX-09 | 
				
					| EME, 
					Infant Flow | 
				
					| Hamilton (夏美顿) Galileo,Raphael (拉斐尔), Alladin HM-12 , 51214 | 
				
					| Hudson 5566 | 
				
					| Infrasonics Adult Star 601T , 60116 | 
				
					| IT-Gambert M-04 、M-16 | 
				
					| Maxtec (迈克思泰克) MAX-12 、MAX-16 | 
				
					| Mercury Medical 10-103-06 | 
				
					| Ohmeda O2 cell 6051-0000-222 | 
				
					| Pacifitech PT-12 | 
				
					| Puritan Bennett (PB泰科) Ventilators 840, 740, 760 4-072214-00 | 
				
					| PPG 
					Biomedical Saracap 504055-001 | 
				
					| Sensidyne SV-12A | 
				
					| Sensor Tec (BPR Medical) ST-12, ST-04 | 
				
					| Siemens (西门子) Servo Ventilators, 900, 900C, 900D, 700, 710, 
					300, V710300 , 9004979 , 6419332 | 
				
					| Space Labs 90514 , Gas Analyzer 015-0132-03 | 
				
					| Taema (天马) YR049700, Cesar 2, Horus | 
				
					| Teledyne R24 , R24MED , Cheiron R-24 , Hill-ROM R-24 | 
				
					| Vascular Technology 103100 | 
				
					| Ventrex V-12A | 
				
					| VTI 
					103100 | 
				
					| Nuova E-17/M | 
				
					|   | 
					
					
					OOM202-1 | 13mV-16mV temperature compensated | <12 seconds  |  > 1 000 000 % O2 
					hours | 3.5mm mono Jack 
					 | Analytical Isolette 
					6735142 | 
				
					| IT-Gambert M-03 | 
				
					| Maxtec (迈克思泰克) MAX-13 | 
				
					|   | OOM202-2 | 13mV-16mV temperature compensated |  <12 seconds |  > 1 000 000 % O2 
					hours | flying leads with 3pin 
					female molex connector  | Catalyst Research MSA (梅思安) MiniOX IA 806572 | 
				
					| IT-Gambert 
					M-09 | 
				
					| Maxtec (迈克思泰克) MAX-14 | 
				
					| Nuova E-09/0 | 
				
					|   | 
					
					
					OOM202-2S |  9mV-11,5mV temperature compensated | <12 seconds  |  >  1 000 000 % O2 
					hours | AMP MATE-N-LOK/2 circuit
					 | City Technology (城市技术) 
					MOX-10 | 
				
					| IT-Gambert M-11 | 
				
					| Stephan Stephanie 
					(斯迪芬妮) | 
			
			
				
					| Measurement range (测量范围) | 0-100% oxygen | 
				
					| Accuracy and repeatability (精度及重复性) | <1% 
					vol.O2 when calibrated at 100% O2 | 
				
					| Zero offset (线性误差) | 0,5% vol.O2 in 100% N2, applied 5 minutes | 
				
					| Linearity error () | <3% 
					realtive | 
				
					| Cross interference (交叉干扰) | <0,5% vol.O2 response to:  10% CO2 balance N2  80% N2O balance N2  7,5% Halothane balance N2  7,5% Isoflurane balance N2  7,5% Enflurane balance N2  9% Sevoflurane balance N2  20% Desflurane balance N2 | 
				
					| Influence of humidity (温度的影响) | -0.03% rel. per % RH at 25°C | 
				
					| Influence of pressure (压力的影响) | proportional to change in oxygen partial pressure | 
				
					| Influence of mechanical shock (机械冲击影响) | <1% 
					relative after fall from 1m | 
				
					| Operating temperature (工作温度) | 0 
					to 50°C | 
				
					| Temperature compensation (温度补偿) | built-in NTC compensation (depends on type) | 
				
					| Effect of temperature compensation (steady state) 
					温度影响补偿(稳态) | between +20°C and +40°C: 3% relative error  between 0°C and +50°C: 5% relative error | 
				
					| Operating humidity (工作湿度) | 0-99% RH non condensing | 
				
					| Long term output drift (长期输出漂移) | <1% 
					vol oxygen per month, typically -<15% relative over lifetime | 
				
					| Storage temperature (存储温度) | -20°C to +50°C | 
				
					| Recommended storage (推荐存储温度) | +5°C to +15°C | 
				
					| Recommended load  (推荐负载) | > 
					10 kOhms | 
				
					| Warm-up time (预热时间) | <30 
					minutes, after replacement of sensor | 
				
					| Weight (重量) | aprrox 
					28 grams, approx 43 grams OOM107 series |